How does AFM work?

The operating principle of atomic force microscopy is depicted in the following schematic:

How does AFM work?

The heart of the AFM lies with the cantilever/tip assembly that interacts with the sample; this assembly is also commonly referred to as the probe. The AFM probe interacts with the substrate through a raster scanning motion. The up/down and side to side motion of the tip as it scans along the surface is monitored through the “beam deflection method”. The beam deflection method consists of a laser that is reflected off the back end of the cantilever and directed towards a position sensitive detector that tracks the vertical and lateral motion of the probe. The deflection sensitivity of these detectors has to be calibrated in terms of how many nanometers of motion correspond to a unit of voltage measured on the detector. Nanosurf instruments provide a straightforward procedure for the calibration of this sensitivity and this procedure is described below in the calibration section.

The probe can also be mounted into a holder with a shaker piezo. The shaker piezo provides the ability to oscillate the probe at a wide range of frequencies (typically 100 Hz to 2 MHz) enabling tapping modes of operation in the AFM. The tapping modes of operation can be performed either in resonant modes (where operation is at or near the resonance frequency of the cantilever) or off-resonance modes (where operation is at a frequency usually far below the cantilever’s resonance frequency).

XYZ motion

Electromagnetic scanners provide highly accurate and precise nanoscale motion in X, Y, and Z at low operation voltage in Nanosurf AFMs. These kinds of scanners provide significant advantages of highly linear motion and the absence of creep over other kinds of scanners such as piezoelectric scanners. The Nanosurf FlexAFM-based systems combine a piezoelectric scanner for Z motion with a flexure-based electromagnetic scanner in X and Y; this configuration provides fast motion in Z with maximum flatness in X and Y, which is optimal for the advanced capabilities offered by these systems.

Atomic force microscopes can be configured either to scan the tip over the sample (in which case the sample is stationary) or scan the sample under the tip (in which case the probe is stationary). All Nanosurf microscopes employ the tip scanning configuration. This configuration provides a significant advantage in terms of flexibility and size of the sample. Tip scanning instruments can accommodate large and unorthodox sample sizes; the only limitation on the sample is that it needs to fit into the instrument! Because the tip is moved and the sample remains stationary, the sample can be almost any size or weight and can still be scanned by the AFM. An example of sample flexibility is shown below with the NaniteAFM system (assembly in typical Nanosurf orange/black at the top) and a custom-built translation/rotation stage to perform roughness measurements on large concave and convex samples.

Nanite AFM in operation

Cantilever/tip (probe) assembly

This assembly consists of a very sharp tip (typical radius of curvature at the end for commercial tips is 5-10 nm) that hangs off the bottom of a long and narrow cantilever. As mentioned previously, the cantilever/tip assembly is also referred to as the probe.

The two most common geometries for AFM cantilevers are rectangular ("diving-board") and triangular. An example of the diving board configuration of the levers is shown in the SEM image below; note the tip hanging off the end.

cantilever, AFM tip images

AFM cantilevers are typically made of either silicon or silicon nitride, where silicon nitride is reserved for softer cantilevers with lower spring constants. The dimensions of the cantilever are very important as they dictate its spring constant or stiffness; this stiffness is fundamental to governing the interaction between the tip and the sample and can result in poor image quality if not chosen carefully. The relationship between the cantilever’s dimensions and spring constant, k, is defined by the equation:

k = Ewt 3 / 4L3,

where w = cantilever width; t = cantilever thickness; L = cantilever length and E = Young’s modulus of the cantilever material. Nominal spring constant values are typically provided by the vendor when buying the probes, but there can be significant variation in the actual values.

Nanosurf provides a straightforward manner of calibrating the spring constants of probes, which is described in the section below.

Deflection sensitivity calibration

This calibration is for the detector sensitivity in order to convert volts measured on the photodetector to nanometers of motion. The calibration is performed by measuring a force curve on an "infinitely stiff" surface such as sapphire. The "infinitely stiff" surface is chosen relative to the cantilever such that the cantilever does not indent into the sample during the force curve measurement. Once the force curve is collected of photodetector signal vs. piezo movement, as shown below, the slope of the repulsive portion of the wall is then calculated. This is the deflection sensitivity.

AFM cantilever calibration schematic

Note that on the Nanosurf Flex-ANA instrument and cantilever calibration options of other product lines this detector sensitivity calibration is automated, where multiple curves are collected and the average detector sensitivity value is calculated.

Spring constant calibration

Calibration of the spring constant of rectangular cantilevers is done via the Sader method on Nanosurf instrumentation and implemented for all current product lines. This method relies on inputting the length and width of the cantilever (these dimensions are provided by the vendor and read from a cantilever list in the AFM software). Generally, a thermal noise spectrum of the cantilever is recorded where the room temperature thermal motion is used to drive the cantilever. A sample thermal tuning spectrum is shown below. A single harmonic oscillator model is used to fit the peak in the thermal spectrum in order to extract the resonance frequency and quality factor. All these parameters are then input into the Sader model for hydrodynamic damping of the cantilever in a given environment, which then calculates the spring constant for the lever.
Alternatively, a frequency sweep can be used to calibrate the spring constant. Here the shaker piezo is used to drive the cantilever.

For spring constant calibration it is important that the cantilever is backed off the surface when these frequency sweeps (either by thermal method or piezo) occur. A lift of at least 100 µm off the surface is recommended.

AFM cantilever thermal tuning

Feedback

The final concept that is important to understanding AFM operation is that of feedback. Feedback and feedback parameters are ubiquitous in our life. For example, temperature is the feedback parameter in a thermostat. The user will set the thermostat to the desired temperature, which is referred to as the "setpoint". As the temperature in the environment changes, there is feedback comparing it with the temperature setpoint so that the heater (or air conditioner) knows when to turn on and off in order to keep the temperature at the desired setpoint value.

Similarly in atomic force microscopes, depending on the different modes, there is a parameter that serves as the setpoint. For example, in static mode the feedback parameter is cantilever deflection, while in the most common form of tapping mode, the cantilever oscillation amplitude is the feedback parameter. The instrument is trying to keep this feedback parameter constant at its setpoint value by adjusting the z piezo to move the cantilever probe up and down. The resulting z piezo movements provide the height information to create the surface topography.

Control of the feedback loop is done through the proportion-integral-derivative control, often referred to as the PID gains. These different gains refer to differences in how the feedback loop adjusts to deviations from the setpoint value, the error signal. For AFM operation, the integral gain is most important and can have a most dramatic effect on the image quality. The proportional gain might provide slight improvement after optimization of the integral gain. The derivative gain is mainly for samples with tall edges. If gains are set too low, the PID loop will not be able to keep the setpoint accurately. If the gains are chosen too high the result will be electrical noise in the image from interference from the feedback; the compensation for a deviation from the setpoint is larger than the error itself or noise gets amplified too strongly.

The other parameters that are important in feedback are the scan rate and the setpoint. If the scan rate is too fast, the PID loop will not have sufficient time to adjust the feedback parameter to its setpoint value and the height calculated from the z piezo movement will deviate from the true topography at slopes and near edges. Very slow scan rates are typically not an issue for the PID loop, but result in long acquisition times that can pose their own challenges such as thermal drift. Optimization of the PID gains and the scan rate are necessary in order to optimize feedback loops. The setpoint affects the interaction force or impuls between probe and sample. A setpoint close to the parameter value out of contact feedback is most gentle for the sample, but tends to slow down the feedback.

See below for an image that was collected with various PID gain settings at the same scan rate. In the red area the image is all electrical noise, because the gains are set too high. The area framed in orange also has some streaks of electrical noise illustrating the same problem. On the bottom, in the blue section, there is poor tracking due to gains being too low. A selected too high scan rate would have a similar appearance. The optimal image and parameter settings are in the green area.

AFM feedback figure